Artis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 11327700

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
April 11, 2022
Recall Number
Z-1246-2022
Quantity
13 units
Firm Location
Malvern, PA

Reason for Recall

if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm

Distribution

US Nationwide distribution.

Lot / Code Info

UDI-DI: 04056869149325 Serial Number: 170026 170305 170307 170309 170310 170023 170304 170025 170051 170312 170027 170303 170053

Root Cause

Device Design

Action Taken

Siemens Medical Solutions USA, Inc. has initiated a Customer Safety Advisory Notice to all affected customers via AX027/21/S on 4/11/22. Siemens will conduct an inspection of the protective earth via Update Instruction AX026/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein.