CK MB Calibrator, Catalogue Number CK2393

This recall has been terminated (originally issued March 25, 2022).

Company
Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
Recall Initiated
March 25, 2022
Terminated
December 26, 2023
Recall Number
Z-1097-2022
Quantity
1,392
Official Source
View on FDA website ↗

Reason for Recall

Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient Samples.

Distribution

Domestic distribution to WV and PR. International distribution to Bangladesh, Bosnia & Herzegovina, Botswana, China, Colombia, Costa Rica, Croatia, Egypt, Fiji, France, Georgia, Germany, Ghana, Greece, Grenada, Hong Kong, India, Iraq, Italy, Lithuania, Malaysia, Maldives, Malta, Mexico, Morocco, Mozambique, Nepal, Nigeria, Pakistan, Philippines, Poland, Puerto Rico, Romania, Russia, Saudi Arabia, South Korea, Switzerland, Thailand, Turkey, UK, Uruguay, Uzbekistan and Vietnam.

Lot / Code Info

GTIN 05055273201451 Lot 4410CK

Root Cause

Under Investigation by firm

Action Taken

Firm notified customers via email on March 25, 2022. The distributors and customers were instructed to discard previous target sheet and perform calibration with the realigned target value provided with the letter. Also, to complete and return the response form 12187-QA to technical.service@randox.com.