RUSCH Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF numbers: a) 112080050 b) 112080055 c) 112080060 d) 112080065 e) 112080070 f) 112080075 g) 112080080 h) 112080085 i) 112080090 j) 112080095 k) 112080100

Company
Teleflex Medical Europe Ltd Ida Business Park Athlone Ireland
Recall Initiated
July 16, 2021
Recall Number
Z-2321-2021
Quantity
2610 units

Reason for Recall

The firm received reports that the endotracheal tube cuff could be inflated while the pilot balloon remained flat.

Distribution

US Nationwide distribution.

Lot / Code Info

a) 112080050, Lot Numbers: KME20K1512, KME20L1145, KME20M2383, KME21B0357 b) 112080055, Lot Numbers: KME20K2579, KME21A2569, KME21C1978 c) 112080060, Lot Numbers: KME20M0616, KME20M2112, KME21A2685 d) 112080065, Lot Numbers: KME20K2581, KME21C1701 e) 112080070, Lot Numbers: KME20K2577, KME20L1985, KME20M1771, KME20M1772, KME20M2856, KME21A2355, KME21B1624, KME21B2490, KME21C1184 f) 112080075, Lot Numbers: KME20K0110 g) 112080080, Lot Numbers: KME20M0377, KME21B2727, KME21C0500 h) 112080085, Lot Numbers: KME20M0909, KME21B0519, KME21C0067 i) 112080090, Lot Numbers: KME20K1808, KME20L2352, KME20M2248 j) 112080095, Lot Numbers: KME20K2310, KME20L1153 k) 112080100 Lot Numbers: KME20K1005, KME20L1144

Root Cause

Process control

Action Taken

The firm initiated the recall by letter on 07/16/2021. The letter explained the issue and requested the return of the product. Distributors were directed to notify their customers.