RecallDepth

PC Application Software c4D as part of the device Catalyst: SP-002 Software PA-003 version 6.1.1 and 6.1.0

Company
C-Rad Positioning Ab Bredgrand 18 Uppsala Sweden
Recall Initiated
August 6, 2020
Terminated
April 15, 2022
Recall Number
Z-2075-2021
Quantity
7

Reason for Recall

PC Application Software c4D not changing Site upon synchronization during the Setup workflow step

Distribution

US Nationwide distribution in the states of IL, KY, MO and WA.

Lot / Code Info

Catalyst: SP-002, Software (part number): PA-003 version 6.1.1 and 6.1.0

Root Cause

Component change control

Action Taken

On August 11, 2020, an e-mail letter containing PA-003-200806-001-01, c4D not changing Site upon synchronization during the Setup workflow step (attached) has been sent personally to our contact persons at the affected sites. In a second step C-RAD upgraded the software at all affected sites to bring back the system to the originally intended state..

More recalls by C-Rad Positioning Ab Bredgrand 18 Uppsala Sweden