Inspire Model 3028, IV Implantable Pulse Generator

This recall is currently active, issued June 21, 2024. It was issued by Inspire Medical Systems Inc..

Company
Inspire Medical Systems Inc.
Recall Initiated
May 16, 2024
Posted
June 21, 2024
Recall Number
Z-2159-2024
Quantity
24 devices
Firm Location
Golden Valley, MN
Official Source
View on FDA website ↗

Reason for Recall

There is a potential manufacturing defect which can lead to electrical leakage in the sensing circuit which causes a system malfunction and requires a revision surgery to replace the IPG,

Distribution

Worldwide distribution - US Nationwide and the country of Germany.

Lot / Code Info

US: UDI/DI 0855728005915, Serial Numbers: AIR324492C, AIR324500C, AIR324506C, AIR324510C, AIR324518C, AIR324523C, AIR324493C, AIR324502C, AIR324507C, AIR324511C, AIR324520C AIR324494C, AIR324503C, AIR324508C, AIR324515C, AIR324521C, AIR324496C, AIR324504C, AIR324509C, AIR324517C, AIR324522C; Europe: UDI/DI 10855728005847, Serial Numbers: AIR324497C, AIR324499C, AIR324505C

Root Cause

Process control

Action Taken

Inspire Medical systems issued an URGENT Medical Device Correction notice to the affected healthcare providers on 05/16/2024 via FedEx overnight. The notice explains the potential problem with the device and risk to the patient and recommends the following actions: We recommend the following actions: " Notify affected patients of this voluntary recall. " Schedule your patient for a monitoring visit to confirm their Inspire therapy is working as intended through waveform and impedance analysis. Monitor patients for reports of variation in stimulation, a lack of therapy effectiveness, or an inability to turn on therapy. " Continue with routine monitoring and conduct waveform and impedance analysis at every routine office visit, as IPG issue may be non-invasively identified with basic waveform and impedance testing. " If abnormal waveforms or impedances are observed, record impedance values at 1.5V, 2.0V, 2.5V, 3.0V and 3.5V. o Let the patient know they will feel 3 short pulses of stimulation. o Some of these stimulation pulses may feel strong. This is normal. " Contact your Inspire representative for additional diagnostic testing and troubleshooting recommendations. For additional questions: please contact your Inspire representative or the Inspire Quality team at ProductRecalls@inspiresleep.com.