Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012, Software Version Number 8.1.100.
This recall is currently active, issued May 24, 2024. It was issued by Philips Medical Systems Nederland B.V. Veenpluis 4-6 Best Netherlands.
- Recall Initiated
- May 16, 2024
- Posted
- May 24, 2024
- Recall Number
- Z-1895-2024
- Quantity
- 1 unit (US)
- Official Source
- View on FDA website ↗
Reason for Recall
When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
Distribution
Domestic: OK International: India, Spain, U.A.E.
Lot / Code Info
Product number: 722012, UDI/DI: 00884838059054; Serial Number: 2138.
Root Cause
Under Investigation by firm
Action Taken
On May 14, 2024 URGENT Medical Device Correction letters were sent to consignees. Actions that should be taken by the customer / user to lower risks for patients: " Circulate this Urgent Medical Device Correction letter to all users so that they are aware of the issue. " Philips recommends discontinuing use of the system for procedures that can be safely postponed to another time, or for procedures that can be performed with another system. " Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure the letter is in a place likely to be seen/viewed. " Please complete and return the attached response form (on page 3) to Philips promptly and no later than 15 days from receipt. Completing this form confirms receipt of the Medical Device Correction letter and understanding of the issue and required actions to be taken. 5. Actions planned by Philips IGT-S to correct the issue: Philips will be re-installing the previous software release in all affected systems. Philips will contact affected customers immediately to schedule a visit to perform this software change (reference FCO72200588). If you need additional information or support concerning this issue, please contact your local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center).