SW Bari Lift & Transfer, Model Number 38060000
This recall is currently active, issued April 18, 2024. It was issued by Raye'S Inc..
- Company
- Raye'S Inc.
- Recall Initiated
- January 25, 2024
- Posted
- April 18, 2024
- Recall Number
- Z-1595-2024
- Quantity
- 1,149 units
- Firm Location
- Hays, KS
- Official Source
- View on FDA website ↗
Reason for Recall
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, UK.
Lot / Code Info
UDI/DI 00845699001040, All serial numbers
Root Cause
Device Design
Action Taken
Sizewise Manufacturing issued an URGENT Voluntary Medical Device Recall to its consignees on 01/29/2024 by letter. The notice explained the issue, potential risk, and requested the discontinuation of use and disposal of the device. The firm sent a follow-up notice on 03/07/2024 asking the distributors to further notify their customers (users of the device).