SP CAL (LIQ), REF IT2692 This product is intended for in vitro diagnostic use with Randox Alpha-1-Antitrypsin, Alpha-1-Acid Glycoprotein, IgA, IgG and IgM assays that require sample predilution.

This recall has been terminated (originally issued November 19, 2018).

Company
Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
Recall Initiated
November 19, 2018
Terminated
April 23, 2020
Recall Number
Z-0660-2019
Quantity
250 kits (32 kits US / 218 kits OUS)
Official Source
View on FDA website ↗

Reason for Recall

Calibrator value sheets have been modified for assigned value of Alpha-1-Acid-Glycoprotein (AGP), IgA and IgG to improve the alignment. IgG measurement reference has also been corrected.

Distribution

US Distribution to states of : IN, NJ, NY, TX and WV and internationally to Ireland.

Lot / Code Info

05055273204049 Lot Number: 415879 Date of Expiry - 28 March 2019

Root Cause

Under Investigation by firm

Action Taken

On November 19, 2018, Randox Laboratories Ltd. issued "Urgent Medical Device Correction" notices and New Calibrator Value sheets (Revised 15 Nov 18 pl) to customers. Customers were advised to take the following action: - Review all remaining stock. Add the notice and replace the calibrator value sheet with the revised document provided. - Discuss the contents of the notice with your Medical Director. Review results generated with the affected batches in line with the clinical profile of the patient. - Retain a copy of the letter with laboratory records and forward a copy of the letter to those who may have received this product. - Return the completed response form via email to technical.services@randox.com. If you have any questions, contact RA Specialist by email at maria.kelly@randox.com.