Optilite Analyser, Product Code: IE700 Product Usage: The Thermo Scientific Indiko Clinical Chemistry Analyzer is a fully automated random access analyzer used to measure a variety of analytes that may be adaptable to the analyzer depending on the reagent used. The Indiko Glucose (H-K) test system, is intended for in vitro diagnostic use in the quantitative determination of the glucose concentration in human plasma on the Indiko analyzer. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemnia, and of pancreatic islet cell carcinoma.

This recall has been terminated (originally issued May 13, 2015).

Company
The Binding Site Group, Ltd.
Recall Initiated
May 13, 2015
Terminated
January 17, 2017
Recall Number
Z-1018-2017
Quantity
129 units
Firm Location
Birmingham United Kingdom
Official Source
View on FDA website ↗

Reason for Recall

Potential risk of the Optilite lid/cover falling suddenly and causing injury when not placed in the fully open position.

Distribution

Worldwide Distribution - US Nationwide in the states of: CA, VA, GA, PA, NJ, KY, NY, AR, NM, NC, AL, GA, MI, and the countries of Canada, China, Australia, Israel, Ireland, UK, Germany, Belgium, Italy, France, Czech republic, Spain, Portugal.

Lot / Code Info

864000290243  864000290832 864000290244  864000290833 864000290330  864000290834 864000290447  864000290836 864000290448  864000290837 864000290449  864000290838 864000290452  864000290844 864000290826  864000290845 864000290827  864000290846 864000290828  864000290847 864000290831  864000290850

Root Cause

Component design/selection

Action Taken

Binding Site sent an Urgent Field Safety Notice letter dated May 13, 2015 to their customers The letter identified the affected product, problem and actions to be taken. Customers are instructed to complete and return the Eback Form that this notice has been received and the actions to be taken have been completed to technical.services@bindingsite.co.uk