Stryker FlowPort II Adapter: Model number: 00CAT00778 The FlowPort II Adapter is intended to connect the FlowPort II Cannulas to commercially available, 4mm arthroscopes in surgical procedures.

This recall has been terminated (originally issued June 3, 2016).

Company
Stryker Corporation
Recall Initiated
May 4, 2016
Posted
June 3, 2016
Terminated
June 3, 2016
Recall Number
Z-1910-2016
Quantity
21 units
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

Complaints were received for the Stryker FlowPort II Adapter, and investigation found that the scope sat in the adapter 180 degrees in the wrong direction

Distribution

GA, UT, MD, NJ, MT

Lot / Code Info

Lot Numbers Affected: 13604

Root Cause

Device Design

Action Taken

A recall letter was not sent as all affected devices have been returned. A ship hold was placed on the device on October 16, 2015.