Microdot Xtra blood glucose test strips

Recall Initiated
July 26, 2018
Recall Number
Z-2357-2019
Quantity
17,784 (USA)

Reason for Recall

The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.

Distribution

Illinois

Lot / Code Info

LOT NUMBERS 7091101, 7091201, 7091801, 7091902, 7100402, 7100403, 7101002, 7101602, 7101702, 7102403, 7103002, 7110602, 7110703, and 7112702.

Root Cause

Nonconforming Material/Component

Action Taken

On July 26, 2018, the firm sent an email to it's US distributor that requested a sub-recall.