Cardiohelp Emergency Drive, Part No. 70104.8002 Product Usage: The Cardiohelp Emergency Drive is used in emergencies to manually drive the disposable if the Cardiohelp-i fails.

This recall has been terminated (originally issued March 28, 2019).

Company
Maquet Cardiovascular Us Sales, Llc
Recall Initiated
February 8, 2019
Posted
March 28, 2019
Terminated
May 14, 2020
Recall Number
Z-0986-2019
Quantity
10
Firm Location
Wayne, NJ
Official Source
View on FDA website ↗

Reason for Recall

Upon servicing the gearwheel in the Emergency Drive the gearwheel may have been re-assembled incorrectly, leading to the inadequate operation of the Emergency Drive.

Distribution

US Nationwide Distribution in the states of California, Florida, Illinois, Kentucky, Michigan, and New York.

Lot / Code Info

Affected serial numbers: 90420359, 90421204, 90421188, 90421279, 90421228, 90421761, 90420904, 90420231, 90420455, 90421304

Root Cause

Process control

Action Taken

Maquet has issued an Urgent Medical Device Recall dated February 26, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that their Emergency Drives had already been inspected and replaced with a loaner Emergency Drive if found to be affected. Affected E-Drives were returned to Getinge for servicing. Upon servicing, a service technician will return the Emergency Drive to your facility. Customers were asked to complete and sign the customer response form to acknowledge receipt of the notification. If you have any questions, please contact your Maquet Cardiopulmonary (MCP)/Getinge representative or call the Maquet Cardiopulmonary (MCP)/Getinge Customer Support at +1 (888) 627-8383 (press option 2, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).