RecallDepth

Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).

This recall has been terminated (originally issued February 11, 2016).

Company
Orthofix, Inc
Recall Initiated
January 8, 2016
Posted
February 11, 2016
Terminated
November 16, 2016
Recall Number
Z-0797-2016
Quantity
36 units
Firm Location
Lewisville, TX
Official Source
View on FDA website ↗

Reason for Recall

There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.

Distribution

Worldwide Distribution -- US, Spain, Italy, and Australia.

Lot / Code Info

BS535418105, BS535419118, BS535420030,  BS535419124, BS562325021, BS535418158, BS535418038, BS535418138, BS535418012, BS535419006, BS535420039,  BS535418127,  BS549024010, BS535419141, BS535418081, BS535418111, BS535418117, BS535420041, BS535418144, BS535420081, BS535419151, BS562325006, BS546825008, BS535418002, BS535418013, BS535418131, BS535420006, BS535420010, BS535420026, BS535420047, BS556219012, BS556219026, BS535418109, BS535420082, BS557717009, and  BS562325023.

Root Cause

Component design/selection

Action Taken

Affected consignees were notified via an Urgent Medical Device Recall Notification letter, dated 1/8/16. The letter identified the affected device and stated the reason for the recall. The recalling firm requests customers to remove the affected devices from their inventory and cease further distribution or use. Customers whom the devices were further distributed to should be notified that the affected devices cannot be used and must be returned to Orthofix per the instructions provided. The attached Acknowledgement Form should be completed and returned. Customers can contact their local Orthofix representative or customer service for further information, replacements, or disposal instructions.