RecallDepth

Cardinal Health Tiny Toes; 11470-010T Infant heel warmer

This recall has been terminated (originally issued December 24, 2015).

Company
Philips Electronics North America Corporation
Recall Initiated
November 20, 2015
Posted
December 24, 2015
Terminated
August 7, 2017
Recall Number
Z-0543-2016
Quantity
7,327,100 individual units
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

Heel warmers may cause first and second degree burns.

Distribution

Moberly, MO

Lot / Code Info

All product in distribution

Root Cause

Labeling design

Action Taken

Cardial Health was notified of the recall via letter on 11/20/2015. Cardinal Health notified all of their customers as of 12/1/2015. Philips instructed Cardinal Health do take the following steps: 1.) Check for and destroy any inventory of Cardinal Health Tiny Toes. 2.) Complete the Distributor Reply Form included in this letter to document this action. Fax to Philips Healthcare at +1 404-855-4900. Please do not return any product to Philips. 3.) Determine which of your customers may have affected product. 4.) Ask your customers to follow the instructions in the section entitled: ACTIONS FOR CARDINAL HEALTH CUSTOMERS/USERS TO TAKE. 5.) As you receive signed Customer Reply Forms from your customers, please forward a running monthly summary to us. Please email the summary to Jennifer Maglio at jennifer.maglio@philips.com. To maintain complete documentation, please retain your customers completed reply forms. Philips may request access to particular reply forms to respond to an audit inquiry. Actions for Cardinal Health Customers to Take: 1.) Discontinue use of Cardinal Health Tiny Toes. 2.) Check for any inventory of Cardinal Health Tiny Toes. 3.) Destroy all products on hand (eg. by activating or puncturing each unit). 4.) Complete and sign the Customer Reply Form provided in this letter or Cardinal Healths version thereof, to notify Cardinal Health of how many Cardinal Health Tiny Toes were destroyed. Submit your signed reply form only after you have completed all steps

More recalls by Philips Electronics North America Corporation

View all recalls by this company →