BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 3.0 mL (Tan) BD Hemogard Additive: K2EDTA (spray dried), 5.4mg Sterile IVD BD, Made in USA BD Vacutainer tubes are used to collect, transport and process blood in a closed system. These tubes are used for testing serum, plasma or whole blood in the clinical laboratory. The BD Vacutainer Plus K2EDTA Tubes may be used for routine immunohematology testing and blood donor screening.
This recall has been terminated (originally issued December 31, 2015).
- Company
- Becton Dickinson & Company
- Recall Initiated
- December 3, 2015
- Posted
- December 31, 2015
- Terminated
- October 19, 2018
- Recall Number
- Z-0570-2016
- Quantity
- 29,815,600 units in total
- Firm Location
- Franklin Lakes, NJ
- Official Source
- View on FDA website ↗
Reason for Recall
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
Distribution
Worldwide Distribution: US (Nationwide) and the countries of: European Union, Peoples Republic of China, Japan and Republic of Korea.
Lot / Code Info
Catalog Reference #367855; Lot numbers 5064684, 5120653, 5154643, 5175930, 5210765, 5237787
Root Cause
Process change control
Action Taken
The firm, BD, issued an "URGENT PRODUCT RECALL" notifications/Recall Response Forms to their customers dated 12/3/2015 via email and/or UPS 2nd Day Mail. The notification identified the reason for the recall; how to identify affected product; and actions to be taken. The customers were instructed customers to immediately review their inventory; complete the enclosed Recall Response Form and fax (1-866-873-0312 it to BD or email it to bd4964@stericycle.com; and return all affected products with the completed Recall Response Form following instructions on the enclosed packing instruction, even if you do not have any affected lot in your inventory. If customers have any questions or require assistance with the return of the recalled product and/or availability of replacement product, they were instructed to contact 1-888-628-0732 between 8AM and 5 PM ET Monday through Friday.