ARCHITECT System Software v5.00 and v5.10, List Number 5F48-22 and 5F48-23; Part Number 7-204717-01 and 7-204717-02 when used with PM Algorithm LLS Brd, part number 7-94255-01 and PM LLS kit, U10 Chip and Ejectors, part number 7-200788-01. The Abbott ARCHITECT System is designed to perform automated: Chemistry tests utilizing photometry and potentiometric technology, and immunoassay tests utilizing CMIA detection technology.
This recall has been terminated (originally issued April 9, 2010).
- Company
- Abbott Laboratories, Inc
- Recall Initiated
- September 11, 2009
- Posted
- April 9, 2010
- Terminated
- January 25, 2011
- Recall Number
- Z-1316-2010
- Quantity
- 11 in the USA; 104 worldwide
- Firm Location
- Irving, TX
- Official Source
- View on FDA website ↗
Reason for Recall
ARCHITECT System Software v5.00 and v5.10 are incompatible with ARCHITECT LLS Board. Software patch re-establishes correct parameters.
Distribution
Worldwide.
Lot / Code Info
List numbers 5F48-22 and 5F48-23, part numbers 7-204717-01 and 7-204717-02 when used with 7-94255-01 and 7-200788-01.
Root Cause
Software design
Action Taken
Technical Service Bulletins dated September 11, 2009 were issued to affected consignees. The correction was issued via the Technical Service Bulletins (TSB) to provide a software patch to re-establish the correct parameters on the LLS board. Direct questions to Abbott Laboratories, Inc. by calling 1-972-518-6000.
More recalls by Abbott Laboratories, Inc
Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479
The Alinity...
Feb 3, 2022
Alinity s System REF 06P16-01. An automated immunoassay analyzer for In Vitro diagnostic use in bl...
Jul 12, 2021
ARCHITECT C System Mixer - Product Usage: The ARCHITECT Clinical Chemistry System is designed to per...
May 1, 2019