ARCHITECT System Software v5.00 and v5.10, List Number 5F48-22 and 5F48-23; Part Number 7-204717-01 and 7-204717-02 when used with PM Algorithm LLS Brd, part number 7-94255-01 and PM LLS kit, U10 Chip and Ejectors, part number 7-200788-01. The Abbott ARCHITECT System is designed to perform automated: Chemistry tests utilizing photometry and potentiometric technology, and immunoassay tests utilizing CMIA detection technology.

This recall has been terminated (originally issued April 9, 2010).

Company
Abbott Laboratories, Inc
Recall Initiated
September 11, 2009
Posted
April 9, 2010
Terminated
January 25, 2011
Recall Number
Z-1316-2010
Quantity
11 in the USA; 104 worldwide
Firm Location
Irving, TX
Official Source
View on FDA website ↗

Reason for Recall

ARCHITECT System Software v5.00 and v5.10 are incompatible with ARCHITECT LLS Board. Software patch re-establishes correct parameters.

Distribution

Worldwide.

Lot / Code Info

List numbers 5F48-22 and 5F48-23, part numbers 7-204717-01 and 7-204717-02 when used with 7-94255-01 and 7-200788-01.

Root Cause

Software design

Action Taken

Technical Service Bulletins dated September 11, 2009 were issued to affected consignees. The correction was issued via the Technical Service Bulletins (TSB) to provide a software patch to re-establish the correct parameters on the LLS board. Direct questions to Abbott Laboratories, Inc. by calling 1-972-518-6000.