DeVilbiss Model 7310PR-D VacuAide Compact Suction Unit. A portable, AC-powered or compressed air-powered suction pump. The device may include a microbial filter.

This recall has been terminated (originally issued February 10, 2009).

Company
Sunrise Medical
Recall Initiated
October 27, 2008
Posted
February 10, 2009
Terminated
September 25, 2012
Recall Number
Z-0863-2009
Quantity
78 units
Firm Location
Somerset, PA
Official Source
View on FDA website ↗

Reason for Recall

Improper suction: Regulator knob does not turn properly which can affect the suction level.

Distribution

Worldwide Distribution: USA, Mexico, Guatemala, Colombia, and Costa Rica.

Lot / Code Info

Model number 7310PR-D. Serial numbers PRD000006, PRD000012, PRD000013, PRD000014, PRD000046, PRD000082, PRD000083, PRD000084, PRD000088, PRD000090, PRD000091, PRD000099, PRD000100, PRD000101, PRD000102, PRD000152, PRD000174, PRD000175, PRD000177, PRD000184, PRD000186, PRD000189, PRD000191, PRD000193, PRD000201, PRD000202, PRD000203, PRD000206, PRD000207, PRD000209, PRD000225, PRD000226, PRD000229, PRD000230, PRD000262, PRD000269, PRD000323, PRD000341, PRD000355, PRD000363, PRD000364, PRD000368, PRD000370, PRD000371, PRD000373, PRD000374, PRD000375, PRD000383, PRD000385, PRD000386, PRD000388, PRD000389, PRD000390, PRD000392, PRD000393, PRD000399, PRD000400, PRD000401, PRD000403, PRD000405, PRD000406, PRD000407, PRD000409, PRD000411, PRD000415, PRD000417, PRD000419, PRD000420, PRD000422, PRD000425, PRD000426, PRD000428, PRD000432, PRD000443, PRD000447, PRD000448, PRD000449, PRD000450, 0425198002000, and 04275195001000

Root Cause

Device Design

Action Taken

DeVilbiss notified dealers and users (through the dealers) to recall the units, by a telephone call October 27, 2008 followed by an Urgent Medical Device Recall mailing November 3, 2008. The letter advised customers what models are affected and how to contact the firm of dealer to have the affected unit replaced. Corrective and preventative action on current manufacturing has been implemented.