RecallDepth

FHC microTargeting Platform DBS Measuring Fixture, a component of the micro-targeting Drive System (Catalog Number 66-FA-SF). The device is a stereotactic instrument used for the placement of recording and stimulating electrodes in the brain.

This recall has been terminated (originally issued December 22, 2008).

Company
Fhc, Inc.
Recall Initiated
October 8, 2008
Posted
December 22, 2008
Terminated
May 2, 2012
Recall Number
Z-0486-2009
Quantity
5 units
Firm Location
Bowdoinham, ME
Official Source
View on FDA website ↗

Reason for Recall

Measuring fixture is incorrectly graduated.

Distribution

Nationwide Distribution including states of CO, FL, NC and NM.

Lot / Code Info

Not coded.

Root Cause

Process control

Action Taken

FHC provided letter notification to FHC Field Representatives on 10/08/08 to hand deliver to customers. FHC Field Representatives will remove incorrectly marked product and provide replacement product. Direct questions to an FHC representative or FHC Quality and Regulatory Affairs at 207-666-5651, ext. 2210.

More recalls by Fhc, Inc.

View all recalls by this company →