RecallDepth

bioMerieux, BacT/ALERT FA Culture Bottles, bioMerieux, Inc., Durham, NC 27704

This recall has been terminated (originally issued January 4, 2008).

Company
Biomerieux, Inc.
Recall Initiated
November 5, 2007
Posted
January 4, 2008
Terminated
December 8, 2009
Recall Number
Z-0457-2008
Quantity
7,025
Firm Location
Durham, NC
Official Source
View on FDA website ↗

Reason for Recall

Duplicate Bottle Indentification -- Three lot numbers of BacT/ALERT FA Culture Bottles labels contained duplicate bottle identification (Bottle ID) barcode numbers. (Lots 1017685 and 1017788 contain the same series of Bottle ID's; Lot 1017693 contains duplicate Bottle ID's within the lot and may share Bottle ID's with both lots 1017685 and 1017788)

Distribution

Worldwide Distribution-USA including countries of Australia, Austria, Bangladesh, Brazil, Bulgaria, Byelorussia, Chile, China, Colombia, Croatia, Denmark, Dijbouti, Estonia, Federation of Russia, Finland, France, Germany, Greece, Guadeloupe, India, Israel, Italy, Korea, Kuwait, Martinique, Mexico, Netherlands, New Caledonia, Norway, Panama, Poland, Portugal, Reunion, Saudi Arabia, Serbia Montenegro, Slovenia, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Vietnam.

Lot / Code Info

Lot Numbers: 1017685, 101793 and 1017788

Root Cause

Process control

Action Taken

bioMerieux notified all Consignees with a Urgent Product Recall Notice letter on 11/05/2007. They were instructed to examine their inventory to determine if they have any remaining inventory of the 3 affected lots. If they have any of these bottles they were instructed to stop using them and destroy all remaining inventory. A Customer Acknowledgement Form was attached to be completed and returned to bioMerieux via fax or mail. There were also instructions attached for the use of Generic Bottle Barcodes and Creating a Unique Bottle ID.