Ellex, Eye Cubed, I 3 System-ABD Ultrasound Version 3.0 with Software Version 3.0 and Biometry A-scan modality, Ellex Innovative Imaging, Sacramento, CA 95827

This recall has been terminated (originally issued December 28, 2007).

Company
Ellex Inc.
Recall Initiated
September 25, 2007
Posted
December 28, 2007
Terminated
September 8, 2008
Recall Number
Z-0148-2008
Quantity
59
Firm Location
Sacramento, CA
Official Source
View on FDA website ↗

Reason for Recall

Incorrect IOL calculations: software behavior with Version 3.0 Eye Cubed Ultrasound systems running Biometry A-scan enabled Version 3.0.0 software can cause incorrect IOL calculations.

Distribution

Worldwide; including Canada and Mexico. U.S.A. distribution included states of AL, AZ, CA, CT, FL, GA, IL, MA, ME, MI, MN, NC, NE, NJ, NM, NY, OH, OR, PA, SC, TN, and TX.

Lot / Code Info

59 systems are affected with serial numbers: V305-00510 V307-18205 V307-18405 V307-20309 V307-20409 V307-20609 V307-20809 V307-20909 V307-21409 V307-19605 V307-15703 V307-16404 V307-16504 V307-19805 V307-17304 V307-18105 V307-19105 v307-14303 v306-09910 v307-13902 v306-05705 v307-17104 v307-17004 v307-14203 v307-14903 v307-17502 v307-15203 v307-18305 v307-15503 v306-04605 v307-18905 v307-16003 v307-18605 v307-18004 v307-16904 V307-21009 v307-14803 V307-20005 v307-17904 v307-19505 v307-12302 v307-21109 V307-21509 v307-11101 v307-19305 v307-21209 v307-17404 v307-18705 v307-18505 v307-16304 V306-07108 v306-09209 v306-04405 V306-06408 v307-19205 V307-20209 v307-19005 v306-12104 v307-19905

Root Cause

Software design

Action Taken

A Global Customer Service Bulletin (GCSB)-Urgent Notification of Software Correction was issued on 25 September 2007, which provided short term workaround directions pending a free software undate promised in the near future.