HUDSON RCI, Variable Concentration, Large Volume Nebulizer, Catalog Number (REF) 1770 and 41770, Rx Only, Teleflex Medical, Research Triangle Park, NC 27709.

This recall has been terminated (originally issued December 28, 2007).

Recall Initiated
August 22, 2007
Posted
December 28, 2007
Terminated
August 13, 2012
Recall Number
Z-0211-2008
Quantity
215,656 units
Firm Location
Durham, NC
Official Source
View on FDA website ↗

Reason for Recall

Decreased in flow output; through the nebulizer -(when used at higher oxygen concentration settings, this decrease in flow, may lower the oxygen concentration inspired by the patient)

Distribution

Worldwide: USA, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, France, Germany, Guatemala, India, Mexico, Panama, Peru, Philippines, South Korea, Thailand, United Arab Emirates, and The United Kingdom.

Lot / Code Info

Product Code/Catalog Number: 1770; Lot Numbers: 49069, 50069, 51069, 52069, 53069, 01079, 02079, 03079, 04079, 06079, 12079, 1279R, 13079, 15079, 16079, 17079, 17079R, 18079, 19079, 19079R and 20079. PRODUCT CODE/CATALOG NUMBER: 41770, Lot Numbers: 50069, 52069, 01079, 06079, 16079, 17079, 17079R, and 18079.

Root Cause

Release of Material/Component prior to receiving test results

Action Taken

Customers were notified by an Urgent Medical Device Recall letter on/about 08/22/2007. They were instructed to return the affected products to Teleflex Medical or acknowledge that the products no longer exist. In addition, each were instructed to forward the letter if they have further distributed into institutional settings or to homecare patients and retrieve relevant product from those locations. This recall is to the consumer/user level via sub-recall from distributors or direct accounts as per the Recalling firm.