Nihon Kohden Electric Stimulator, Model number: MS-210BK (Optional accessories of MEB-2200A -Neuropack Evoked Potential and EMG Measuring System)

Company
Nihon Kohden America Inc
Recall Initiated
August 23, 2006
Posted
November 28, 2006
Terminated
May 23, 2008
Recall Number
Z-0218-2007
Quantity
97 units in the U.S.
Firm Location
Foothill Ranch, CA

Reason for Recall

When the device is subject to physical shock, such as a drop to the floor, it may output a different value from the preset value. This may result in a superficial burn on the patient if the stimulation is done for a long time.

Distribution

Nationwide.

Lot / Code Info

Any serial numbers up to 02160

Root Cause

Other

Action Taken

All customers will be called by phone, then faxed the Product Notification starting 09/01/2006 Customers were instructed to indicate the serial numbers on each of your MS-210BK stimulators and fax the MS-210BK PRODUCT RECALLREPLACEMENT letter back to Nihon Kohden at: 949-580-15550 or email to customerservice@nkusa.com. Upon receipt of this letter, Nihon Kohden''s customer service department will send a replacement unit.