Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required); Models 801763.

This recall has been terminated (originally issued December 27, 2006).

Company
Terumo Cardiovascular Systems Corp
Recall Initiated
May 1, 2004
Posted
December 27, 2006
Terminated
July 14, 2009
Recall Number
Z-0314-2007
Quantity
12
Firm Location
Ann Arbor, MI
Official Source
View on FDA website ↗

Reason for Recall

The gas system may fail calibration prior to use due to incorrect gas system flowmeter software.

Distribution

Worldwide distribution, including USA, Belgium, Germany, Japan, Korea and Republic of China.

Lot / Code Info

Serial numbers 0093 and 0104 through 0114.

Root Cause

Process control

Action Taken

U.S. consignees were visited and corrections made between May 2004 and November 2005. Foreign affiliates were instructed to make corrections on international units, but the firm has not received confirmation yet that these units have been corrected.