Hill-Rom 100 Low Bed (GPAC) AC Powered adjustable hospital bed

This recall has been terminated (originally issued November 13, 2015).

Company
Hill-Rom, Inc.
Recall Initiated
October 7, 2015
Posted
November 13, 2015
Terminated
February 10, 2017
Recall Number
Z-0274-2016
Quantity
12251 units
Firm Location
Batesville, IN
Official Source
View on FDA website ↗

Reason for Recall

Hill-Rom is voluntarily recalling the Hill-Rom 100 Low Bed due to complaints of patient injury when exiting the beds set at the lowest position setting.

Distribution

Worldwide Distribution - USA (nationwide) to the states of : AL, AR, AZ. CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI and WV., and to the countries of : Canada and France.

Lot / Code Info

Product # P3930A; Dates of Manufacture 5-DEC-2007 to 14-SEP-2015; Serial numbers from I325AR0038 thru Q257AR2299

Root Cause

Device Design

Action Taken

Hill-Rom initiated Phase 1 of the recall on October 2, 2015, which included notices of the recall being issued to customers via certified mail, which included safety information on how to properly use the bed, and instructions to forward the information on if devices have been transferred. Phase 2 for the recall will include the distribution of caps for the edges of the bed to be placed on by the customers. Customers may contact Hill-Rom Technical Support at 800-445-3720.