RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

This recall has been terminated (originally issued May 27, 2015).

Company
Raysearch Laboratories Ab Sveavaegen 9 Stockholm Sweden
Recall Initiated
May 8, 2015
Posted
May 27, 2015
Terminated
April 11, 2017
Recall Number
Z-1679-2015
Quantity
Domestic: 539 units
Official Source
View on FDA website ↗

Reason for Recall

RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing.

Distribution

US Nationwide Distribution

Lot / Code Info

Software Version (Build numbers) 2.5.1.89, 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, or 4.0.3.4

Root Cause

Software Design Change

Action Taken

RaySearch sent a Device Correction letter dated May 8, 2015 via e-mailed to their customers. The letter identified the affected product, problem, actions to be taken by users and contact information.