RecallDepth

Medtronic DBS Extension Kit for Deep Brain Stimulation, Models 37085 and 37086. Sterile

This recall has been terminated (originally issued May 15, 2015).

Company
Medtronic Neuromodulation
Recall Initiated
April 16, 2015
Posted
May 15, 2015
Terminated
July 22, 2016
Recall Number
Z-1617-2015
Quantity
86,851 (US 37,933 / OUS 48,918)
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Medtronic is providing Important information regarding extension handling during implant procedure after receiving reports of high impedances. This applies to extension Models 37085 and 37086 which can be used with the following implantable Neurostimulators: Activa PC (Model 37601), Activa RC (Model 37612), and the Activa SC (Model 37603).

Distribution

worldwide

Lot / Code Info

all Serial numbers

Root Cause

Under Investigation by firm

Action Taken

Medtronic began notifying consignees with an Urgent Medical Device Safety Notification letter beginning April 16, 2015. Consignees will be notified by mail and/or in person, by visit of a Medtronic field representative. The letter described the issue and provided recommendations of following the instructions defined within the Neurostimulator implant manual. A Physician Reply Form was asked to be returned via fax at 800-897-3899 or email to neuro.quality@medtronic.com. Consignees with questions can contact Medtronic Neuromodulation Technical Services at 1-800-707-0933 weekdays 7am - 6pm CST.

More recalls by Medtronic Neuromodulation

View all recalls by this company →