CPR Shield With One Way Valve and Barrier Filter, Reorder No. 4921 --- Dynarex label, 100 per case --- Manufactured for: Dynarex Corporation - Made in China. Used as a physical barrier for mouth to mouth resuscitation.

Company
Dynarex Corporation
Recall Initiated
July 2, 2014
Posted
July 25, 2014
Terminated
April 20, 2017
Recall Number
Z-2117-2014
Quantity
Domestic: 208 cases; Foreign: 42 cases
Firm Location
Orangeburg, NY

Reason for Recall

"THIS SIDE UP" is on the wrong side of the product. This error poses a potential health hazard if the wrong labeling is followed by caregivers.

Distribution

Worldwide Distribution -- USA, Canada, and Mexico.

Lot / Code Info

Lot Numbers 28619 & 28621

Root Cause

Employee error

Action Taken

Voluntary Device Recall letters (dated 7/01/14) and recall verification forms were sent to customers via US Postal First Class Mail on 7/02/14. The letters instructed customers to check their inventory for the specific lots, discontinue distributing them, and quarantine them immediately. Customers were to promptly contact Dynarex Credit & Return Representative Avi Celnick at 1-845-365-8200 ext. 6644 to arrange for product return and obtain an RGA #. For questions, please call Dynarex's toll free number 1-888-396-2739.