CPR Shield With One Way Valve and Barrier Filter, Reorder No. 4921 --- Dynarex label, 100 per case --- Manufactured for: Dynarex Corporation - Made in China. Used as a physical barrier for mouth to mouth resuscitation.

This recall has been terminated (originally issued July 25, 2014).

Company
Dynarex Corporation
Recall Initiated
July 2, 2014
Posted
July 25, 2014
Terminated
April 20, 2017
Recall Number
Z-2117-2014
Quantity
Domestic: 208 cases; Foreign: 42 cases
Firm Location
Orangeburg, NY
Official Source
View on FDA website ↗

Reason for Recall

"THIS SIDE UP" is on the wrong side of the product. This error poses a potential health hazard if the wrong labeling is followed by caregivers.

Distribution

Worldwide Distribution -- USA, Canada, and Mexico.

Lot / Code Info

Lot Numbers 28619 & 28621

Root Cause

Employee error

Action Taken

Voluntary Device Recall letters (dated 7/01/14) and recall verification forms were sent to customers via US Postal First Class Mail on 7/02/14. The letters instructed customers to check their inventory for the specific lots, discontinue distributing them, and quarantine them immediately. Customers were to promptly contact Dynarex Credit & Return Representative Avi Celnick at 1-845-365-8200 ext. 6644 to arrange for product return and obtain an RGA #. For questions, please call Dynarex's toll free number 1-888-396-2739.