Stryker Orthopaedics Triathlon X3 Tibial Bearing Insert - CR REF 5530-G-209 (component of the Triathlon Total Knee System) for fixation to the tibial baseplates via a locking wire mechanism.

This recall has been terminated (originally issued August 14, 2015).

Company
Stryker Howmedica Osteonics Corp.
Recall Initiated
May 29, 2015
Posted
August 14, 2015
Terminated
September 26, 2016
Recall Number
Z-2389-2015
Quantity
10 units
Firm Location
Mahwah, NJ
Official Source
View on FDA website ↗

Reason for Recall

Stryker has received a report which identified that a Triathlon Tibial Bearing Insert CS X3 Size #2 9mm (5531-G-209) was reported to be in a pack that was labelled as a CR insert (5530-G-209).

Distribution

OH, SC, RI, WV

Lot / Code Info

Catalog number 5530-G-209   Lot Code LEF064

Root Cause

Labeling mix-ups

Action Taken

Stryker Orthopaedics sent a recall notification via email on 5/29/2015 to their local branches. Notification letters and Product Accountability Forms were sent to Stryker branches/agencies with return receipts on 7/2/2015.