RenLane Renal Denervation Catheter; Cat No.: D135601. Used in adults patients (>18 years) with drug resistant hypertension to denervate the renal arteries to reduce blood pressure.

This recall has been terminated (originally issued August 18, 2015).

Company
Cordis Corporation
Recall Initiated
April 15, 2014
Posted
August 18, 2015
Terminated
August 26, 2015
Recall Number
Z-2405-2015
Quantity
98 units
Firm Location
Miami Lakes, FL
Official Source
View on FDA website ↗

Reason for Recall

Potential Damage may occur to the helical tip of the RENLANE Renal Denervation Catheter, if the hemostasis valve of the arterial access device is not fully open before insertion or withdrawal of the catheter.

Distribution

Distributed in Germany only.

Lot / Code Info

Model Number: D135601 Lots #'s 15958389LA, 15959088LA, 15959090LA, 15968753LA, 15968806LA, 15968810LA, and 15968813LA, with Exp Date 8/31/2014 Lots #'s 15976735L, 15976820L, 15976823L, 15978174L, 15978179L, 159784180L, and 15978181L with Exp Date 9/30/2014

Root Cause

Labeling Change Control

Action Taken

Cordis sent an Urgent Field Safety Notice dated April 15, 2014 providing additional labeling to their consignees in Germany only, with a follow-up letter on August 4, 2014. Cordis explained When deploying the catheter via a compatible arterial access device, if a Tuohy-Borst type adjustable hemostasis valve is used, ensure the valve is fully open before insertion or withdrawal of the catheter. This is especially important, since the helical tip at the distal end of the RENLANE Renal Denervation Catheter has a larger diameter compared to the proximal catheter shaft. Although the event was not a Removal, all lots involved have expired. The product is no longer commercialized, and was never commercialized beyond the 10 customers in Germany.