HeartWare Ventricular Assist System (HeartWare Controller) Product Usage: For use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HeartWare System is designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter.

This recall has been terminated (originally issued February 25, 2015).

Company
Heartware, Inc.
Recall Initiated
January 16, 2015
Posted
February 25, 2015
Terminated
February 5, 2020
Recall Number
Z-1131-2015
Quantity
4,845
Firm Location
Miami Lakes, FL
Official Source
View on FDA website ↗

Reason for Recall

The affected clinical trial Controllers exhibit a higher susceptibility to ESD than newer commercial Controllers.

Distribution

Worldwide Distribution - US Nationwide in the states of: AL, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NY, OH, OR, PA, TX, UT, VA, WA and District of Columbia and countries of: Canada, Poland, Australia, Austria, Italy, DE, New Zealand, South Africa, France, Switzerland, Germany, Brazil, UK, Belgium, Netherlands, Israel, Malaysia, Turkey, Greece, Japan, Singapore, Poland, Lithuania, Sweden, Norway, Finland, and India.

Lot / Code Info

Product Codes- US product: Model No.: 1400, 1401US; International product: Model No.: 1400, 1401XX ('XX' represents country designation) Serial No.: CON000001 through CON005472

Root Cause

Device Design

Action Taken

HeartWare sent an URGENT MEDICAL DEVICE RECALL dated January 16, 2015 asking clinicians to quarantine the affected controllers and complete and return the acknowledgement form. HeartWare also asks clinicians to identify patients with affected controllers and review the applicable risks with the patient and as soon as reasonably possible if medically advisable, exchange the affected controller under medical supervision with a new controller. HeartWare also asks that for each patient using an affected controller as their backup to contact the patient and arrange to have the backup controller replaced with a new controller. Lastly, HeartWare asks the firm to complete the patient tracking form and return all quarantined controllers to HeartWare along with the completion form. Should you have any questions or concerns, please contact your local HeartWare representative. Our 24-Hour Clinical Support personnel are also available at 1-888-494-6365.