PATH(R) THREAD CUP ADAPTOR, REF 2007-0013, Wright Medical Technology, Inc. Orthopaedic surgical instrument.
- Company
- Microport Orthopedics Inc.
- Recall Initiated
- April 30, 2014
- Posted
- June 4, 2014
- Terminated
- December 1, 2014
- Recall Number
- Z-1712-2014
- Quantity
- 140 units
- Firm Location
- Arlington, TN
Reason for Recall
Difficulty removing the cup adaptor from the impaction handle during surgery
Distribution
Nationwide Distribution.
Lot / Code Info
Lot Number(s): 1153896, 1279714, 1288604, and 1314793
Root Cause
Device Design
Action Taken
Firm has initiated the recall by email on 04/28/2014 to all distributor's offices, including international, then followed with written communication on 05/02/2014 via Fed Ex P1 delivery to all United States consignees. Follow-up to non-responding consignees will commence 30 days after initiation.