Synthes Trauma Nail System. The devices are indication for bone fixation.

Company
Synthes Usa (Hq), Inc.
Recall Initiated
November 2, 2012
Posted
August 26, 2013
Terminated
September 27, 2016
Recall Number
Z-2058-2013
Quantity
1,011,783
Firm Location
West Chester, PA

Reason for Recall

Recall was initiated due to the possibility that the outer pouch was compromised. The product is packaged in two pouches; a sterile, internal pouch within an external pouch. It is the outer, external pouch that may be compromised.

Distribution

Nationwide distribution.

Lot / Code Info

Trauma Ex Nails, all lot numbers up to and including lot # 7072305 and Trochanteric Fixation Nails, all lot numbers up to and including lot # 7060897

Root Cause

Packaging change control

Action Taken

The firm initiated their recall of this product on November 2, 2012 by sending an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter to all consignees. The letter explained the problem and gave instructions for consignees to examine their inventory and urged consignees to return the identified devices. The letter stated If returning the product would adversely impact their ability to provide necessary medical care to patients, they should re-sterilize the product per Pre-Vacuum Sterilization instructions contained in the Instruction for Use.