Synthes Titanium Midface Distractor, MR Conditional, Part number 487.982. For temporary stabilization and gradual lengthening of the cranial or midfacial bones.

Company
Synthes Usa (Hq), Inc.
Recall Initiated
January 11, 2013
Posted
March 12, 2013
Terminated
August 31, 2015
Recall Number
Z-0942-2013
Quantity
596
Firm Location
West Chester, PA

Reason for Recall

There is the potential for the anterior footplates to not fit onto the Titanium Midface Distractor Assembly as the tab of some distractors may be oversized.

Distribution

Worldwide Distribution -- USA, including the states of CA, LA, MA, NY, and PA, and the countries of Canada and Switzerland.

Lot / Code Info

All Lots

Root Cause

Employee error

Action Taken

The firm initiated their recall of this product on January 11, 2013 by sending an Urgent Notice: Medical Device Recall letter to all consignees. The letter identified the affected product and the problem. Also, the firm requested consignees to examine their inventory for products(s) with the specified part number, remove them from use, and return them to Synthes. Customers were to follow the instructions provided on what to do if they have or do not have any identified affected product. Questions should be directed to 610-719-5450 or FieldAction@synthes.com.