Becton, Dickinson and Company, BBL Crystal Enteric/Nonfermenter ID Kit, Catalog # 24500, cartons of 20 kits, In-vitro diagnostic reagents for microbiological testing. The BBL Crystal Enteric/Nonfermentor (E/NF) identification (ID) System is for the identification of aerobic gram-negative bacteria that belong to the family Enterobacteriaceae as well as some of the more frequently isolated glucose fermenting and nonfermenting gram-negative bacilli.

This recall has been terminated (originally issued January 20, 2013).

Company
Becton Dickinson & Co. Bd Diagnostic Systems
Recall Initiated
December 13, 2012
Posted
January 20, 2013
Terminated
June 3, 2013
Recall Number
Z-0699-2013
Quantity
67,820 units
Firm Location
Sparks, MD
Official Source
View on FDA website ↗

Reason for Recall

In-vitro diagnostic reagents for microbiological testing may be reversed in well positions in the test kit and when utilized may lead to bacteria misidentification.

Distribution

Worldwide Distribution-USA including the states of MA, TN, KY, IN, GA, NC, TX, NY, FL, AZ, WA, NJ, MI, and CA, and the countries of Uruguay, Taiwan, Singapore, Australia, South Korea, Mexico, Hong Kong, Guatemala, Colombia, Chile, Canada, India, Japan, Brazil, China and Belgium.

Lot / Code Info

Lot Number/ Exp Date 2086315 February 20, 2013 2101435 February 20, 2013 2104316 February 27, 2013 2129460 February 27, 2013 2129487 March 13, 2013 2163211 February 27, 2013 2163213 March 20, 2013 2163215 March 27, 2013

Root Cause

Mixed-up of materials/components

Action Taken

Becton Dickinson initiated the recall by letter to direct accounts on 12/13/2012. Ex-US BD sites were notified via email, distributors were notified by fax and customer letters were delivered by UPS ground. The notification advised that the recalled kit contained citrate (CIT} and malonic acid (MLO} reagents in the reversed prong/well positions. Customers were advised to discontinue use of identified lot numbers and discard any remaining packages for replacements, and also to return the enclosed reply form. The recalling firm requested distributors to submit their customer list for direct BD notification to all end users. Contact the BD Customer Service Department at 1-800-638-8663 if further assistance is required.