The Instructions For Use for the following: Stryker TRIO+ 4.5 x 45 mm STANDARD POST SCREW Stryker TRIO+ 4.5 x 50 mm STANDARD POST SCREW Stryker TRIO+ 4.5 x 55 mm STANDARD POST SCREW Stryker TRIO+ 5.5 x 45 mm STANDARD POST SCREW Stryker TRIO+ 5.5 x 60 mm STANDARD POST SCREW Manufactured by: STRYKER Spine SAS, Z.I. Marticot - 33610 - Cestas-France +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct., Allendale, NJ 07401-1677 USA +1-201-750-8000. The system is intended for posterior, non-cervical pedicle and non-pedicle pedicle fixation of the spine. It is indicated for: degenerative disc disease (DDD), spondylolisthesis, fracture, dislocation, spinal stenosis, scoliosis kyphosis, lordosis, spinal tumor, pseudarthrosis, and failed previous fusion.

Company
Stryker Spine
Recall Initiated
April 25, 2012
Posted
August 22, 2012
Terminated
February 27, 2013
Recall Number
Z-2235-2012
Quantity
40 units.
Firm Location
Allendale, NJ

Reason for Recall

The front cover of Instructions For Use (IFU) #NOLI135B03Rev02 was improperly marked with a "Sterile" logo in the lower right corner. The products these IFUs were packed with are sold as "Non-Sterile" and require sterilization before use in surgery.

Distribution

Distributed in Atlanta, Georgia.

Lot / Code Info

Code information listed in the following format: Manufacturer Part Number/Prodcut Description/Corresponding IFU Part Number/Affected Lot #s are as follows:  48900445/Trio Standard Post Screw 5.5x55mm/#NOLI135B03Rev02/57281 and 57201  48900450/Trio Standard Post Screw 5.5x60mm/#NOLI135B03Rev02/57197   48900455/Trio Standard Post Screw 4.5x45mm/#NOLI135B03Rev02/57198  48900555/Trio Standard Post Screw 4.5x50mm/#NOLI135B03Rev02/56599  48900560/Trio Standard Post Screw 4.5x55mm/#NOLI135B03Rev02/57192

Root Cause

Labeling mix-ups

Action Taken

Stryker sent an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letter to affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Customer Response Form was attached for customers to complete and return via fax to 201-760-8370. Contact the firm at 201-760-8298 for questions regarding this notice.