Stryker AVS Navigator Trials 6mm-15mm Manufactured by: Stryker Spine SAS Z.1 Marticot-33610-Cestas -France +33 (0) 5.57.97.06.30 http;//www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Court. Allendale NJ 07401-1677 USA +1-201-780-8000 Cestas France 33610. Intended Use: The trials are designed and intended to measure the height of the disc space intra-operatively.

Company
Stryker Spine
Recall Initiated
January 13, 2012
Posted
July 19, 2012
Terminated
July 29, 2013
Recall Number
Z-2033-2012
Quantity
273 units
Firm Location
Allendale, NJ

Reason for Recall

Beginning in February 2011, Stryker Spine began receiving reports related to the shaft of the AVS Navigator trial failing at the junction of the trial head.

Distribution

Nationwide.

Lot / Code Info

Class I; 510(k) Exempt Description: AVS¿ Navigator Trials. Size 6mm -15mm. Catalog Number: 48392206,48392207,48392208,48392209,48392210, 48392211,48392212,48392213,48392214,48392215. Lot Code: 104864,104866, 104867, 104868, 104869, 104870, 10E331, 104871, 104872, 104873, 104874, 104875.

Root Cause

Device Design

Action Taken

Stryker sent an "URGENT PRODUCT RECALL" letter to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Customer Response Form was included for customers to complete and return via fax to Regulatory Compliance at 201-760-8370. Contact the firm at 201-760-8298 for questions regarding this recall.