VARIANT B-thalassemia Short Program Reorder Pack, 500 Tests, For the separation and determination of HbA and HbF and as an aid in the identification of abnormal hemoglobins in whole blood. IVD, Made in United States. Contents: 1. Elution Buffer 1, 3 x 1900 mL sodium phosphate buffer 2. Elution Buffer 2, 2 x 1800 mL sodium phosphate buffer 3. Wash/Diluent solution, 2 x 1800 mL deionized water 4. Analytical cartridge, 1 cation exchange cartridge 5. 1 Calibrator/Diluent set, 1 Lypholized human red blood cells hemolysate with preservative; deionized water with preservative 6. 1 CD Rom - CD R with program parameters 7 Sample vials, 100 x 1.5 mL. Sample vials with pierceable cap 8. Whole Blood Primer, 10 x 1.0 mL (3), Lypholized human red blood cells hemolysate with preservative; 9. Instruction Manual Bio-Rad Laboratories, Hercules, CA 94547. Percentage determination of hemoglobin A1c in human whole blood using ion-exchange high=performance liquid chromatography (HPLC)

Company
Bio-Rad Laboratories Inc
Recall Initiated
September 12, 2011
Posted
October 7, 2011
Terminated
May 11, 2012
Recall Number
Z-0024-2012
Quantity
491 lot 70210674; 298 lot 70210954; 136 lot 70010957
Firm Location
Hercules, CA

Reason for Recall

A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.

Distribution

Nationwide and to New Zealand, Canada, France, England, China, India, Mexico, Italy, Brazil, Australia, Singapore, Thailand, SriLanka, hong Kong, Sweden, Germany, Peru, Jamaica, Uruguay, Columbia, Argentina, Chile, Paraguay and Korea.

Lot / Code Info

Model number: 270-2154; Lot number 70210674, expiry 1/13/13. lot number 70210954, expiry 2/28/13. Both contain Buffer lot numbers AA10685/AA10686, expiry 3/31/13. Model number : 270-0002; Lot number 70010957, expiry 2/28/13. Contains Buffer lot numbers AA10685/AA10686, expiry 3/31/13.

Root Cause

Process control

Action Taken

Bio-Rad sent an Urgent Medical Device Correction letter on September 12, 2011, to all affected subsidiaries and customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discard all affected product in accordance with all local, state and federal disposal procedures. If customers distributed the product further they should direct their customers to discard all unused product. Customers were asked to please translate (if necessary) the attached "Customer Medical Device Correction Response Form," customers should add their local contact information and distribute to all affected customers in their region by mail or fax. Compile tracking documentation (e.g. Excel spreadsheet) for all customers who received the Medical Device Correction Letter. Complete the "SUBSIDIARY MEDICAL DEVICE CORRECTION RESPONSE FORM" and fax to Bio-Rad Regulatory Affairs at (510) 741-3954, or e-mail PDF to CSD_RA/Hercules/US/BIO-RAD. Customers would be issued credits for the affected products based on their request. For questions call (510) 741-3954.