Ratcheting Torque Limiting Handle, Catalog Number 52-1012, Orthofix Spinal Implants, Firebird Spinal Fixation System, Firebird Spinal Fixation System DDD Instrument Case. Product is intended for use with the Set Screw Driver (52-1061) and Adjustable Counter Torque Wrench (52-1265) to perform the final tightening of set screws in a spinal implant construct.

This recall has been terminated (originally issued September 7, 2011).

Company
Orthofix, Inc
Recall Initiated
April 1, 2011
Posted
September 7, 2011
Terminated
March 23, 2012
Recall Number
Z-3162-2011
Quantity
287 units
Firm Location
Lewisville, TX
Official Source
View on FDA website ↗

Reason for Recall

Upon reaching torque limit, there is a possibility the product may jam or exhibit a ratchet direction switch from forward to reverse.

Distribution

Worldwide Distribution-USA (nationwide) and Puerto Rico and the countries of Australia and Germany.

Lot / Code Info

All serial numbers.

Root Cause

Device Design

Action Taken

Orthofix, Inc. sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated April 1, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to identify and remove the affected product from use. Customers were advised to use the combination of the Ratching R Handle and the Set Screw Driver if they experience problems with the Ratching Torque Limiting Handle. A Tracking and Verification form was attached for customers to complete and return to the Orthofix Inc. Regulatory Affairs Department via fax at 214-937-3322. Contact your Account Services Representative at 888-298-5700 for questions regarding the removal and return of the affected product.