3.5MM LOCKING SCREW, T15 HEXALOBE, SELF TAPPING, 34MM The OrthoPediatrics Proximal Femur Plating System combines implants and instruments in one convenient system. This system offers the advantages of the osteotomy blade plates and cannulated instrumentation. Osteotomy plates provide ease of reduction and good rotational stability while maintaining bone stock. Cannulated instruments work over a guide wire for precise placement and safety. The subject device is a Bone Screw designed for use with this system.

Company
Orthopediatrics Corp
Recall Initiated
September 26, 2013
Posted
November 14, 2013
Terminated
December 26, 2013
Recall Number
Z-0276-2014
Quantity
7 units shipped
Firm Location
Warsaw, IN

Reason for Recall

The firm identified, through complaints that the 3.5MM LOCKING SCREW packaging material was marked as a 3.5mm screw and the part number was correct but that the description indicated a 7.0mm screw. Upon investigation, it has been determined that a lot number 009U2 was packaged and mislabeled internally by OrthoPediatrics.

Distribution

Worldwide Distribution - US including Michigan and the countries of Ireland and Australia

Lot / Code Info

Part Number:. 00-0903-2634 and lot number 009U2

Root Cause

Labeling mix-ups

Action Taken

OrthoPediatrics sent an Urgent Medical Device Recall letter dated September 26, 2013, to all affected customers.The letter identified the product the problem and the actions needed to be taken by the customer The sales representatives and distributors who have been identified via shipping records as having received one or more discrepant devices affected by this Recall were requested to quarantine the devices identified in this Recall by telephone on 18-September-2013. If you or any of your customers have any of these devices in the original packaging please take possession of them and quarantine them in preparation for returning them to OrthoPedlatrics. They are misbranded. If the Screw has already been removed from the original packaging and the original packaging has been discarded and the Screw placed in a sterilization tray, it need not be returned. For further questions please call ( 574) 268-6379.