Nova Max Glucose Test Strips (10 Count Vials) Catalog Number: 43604 These strips are kitted into Nova Max Glucose Test Strip System Kits 43435 and 45507 Product Usage: The Nova Max Glucose Test Strips are for use with the Nova Max Family of Glucose Monitors (Nova Max, Nova Max Plus, Nova Max Link, BD Logic and Paradigm Link). These monitors are intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. They are intended for use by people with diabetes mellitus in the home and by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control

Company
Nova Biomedical Corporation
Recall Initiated
July 15, 2013
Posted
August 9, 2013
Terminated
August 1, 2017
Recall Number
Z-1901-2013
Quantity
*Kitted into System Kits 43435 and 45507
Firm Location
Waltham, MA

Reason for Recall

Reports of false abnormally high glucose readings from test strips

Distribution

Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua

Lot / Code Info

Lot #/Exp.Date: 1020212032 02/28/2014 1020212153 06/30/2014 1020212207 07/31/2014 1020212291 10/31/2014 1020213074 03/31/2015 1020213109 04/30/2015 1020412255 09/30/2014 1020412341 12/31/2014 1020512087 03/31/2014 1020811350 12/31/2013 102121304 02/28/2015

Root Cause

Material/Component Contamination

Action Taken

Nova Biomedical U.S.A issued an Urgent Medical Device Voluntary Recall Letter dated July 19, 2013 via registered/certified letters to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately discontinue use of the affected product; contact Nova Diabetes Care Customer Service for replacement strips; do not stop testing blood glucose while waiting for replacement strips; if affected strips must be used follow precautions provided in notification letter. For questions call Nova Diabetes Care Customer Service at 1-800-681-7390. .