BD BBL Taxo XV Factor Strips, Catalog number 231104, in shelf pack vials, fifty strips/vial labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152 USA 800-638-8663, www.bd.com/ds***

This recall has been terminated (originally issued December 13, 2012).

Company
Becton Dickinson & Co. Bd Diagnostic Systems
Recall Initiated
October 22, 2012
Posted
December 13, 2012
Terminated
January 24, 2013
Recall Number
Z-0523-2013
Quantity
650 vials
Firm Location
Sparks, MD
Official Source
View on FDA website ↗

Reason for Recall

In vitro diagnostic test kit was not manufactured according to specifications and may cause false negatives or misidentification of Hemophilus species present in patient specimens.

Distribution

Worldwide Distribution - USA (nationwide) to international customers in Belgium, India, Japan, South Korea, Thailand and Mexico

Lot / Code Info

Lot number 2089253

Root Cause

Employee error

Action Taken

BD Diagnostic Systems sent an Urgent Product Recall letter dated October 22, 2012, to all affected customers.. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to discontinue distribution and discard affected inventory for credit. . Customers were requested to return a response form by fax to 410-316-4258. For questions customers were instructed to contact BD Customer Service Department at 1-800-638-8663. For assistance with replacements, customers were instructed to call 1-800-675-0908. For questions regarding this recall call 410-316-4258.