Invacare 600 lb. Bariatric Bed, Head motor extension cable used on component numbers BAR5490 and BAR5490IVC. These components are part of the following Bed Model Number(s): BAR600, BARPKGCA, BAR600IVC, BARPKGIVC-1633, and BARPKG.

This recall has been terminated (originally issued November 14, 2011).

Company
Invacare Corporation
Recall Initiated
June 23, 2011
Posted
November 14, 2011
Terminated
July 2, 2013
Recall Number
Z-0187-2012
Quantity
35,747
Firm Location
Elyria, OH
Official Source
View on FDA website ↗

Reason for Recall

Invacare Corporation decided to recall the products because risk analysis reveals that miskeying or disconnection of the head motor extension cable during bed set-up may result in product damage, or in rare instances, the potential for fire.

Distribution

Distribution Nationwide and to United Arab Emirates, Dominican Republic, Egypt, Guam, Mexico, & Virgin Islands.

Lot / Code Info

Model Number(s): BAR600, BARPKGCA, BAR600IVC, BARPKGIVC-1633, and BARPKG. 35747 serial numbers available from Invacare.

Root Cause

Labeling design

Action Taken

On 9/19/11, Invacare sent Recall Notificaiton Letters to their customers. They instructed their customers to quarantine the product, complete and return a response form.