Haemonetics cardioPAT Cardiovascular - autotransfusion System Catalog Number: 02050-US

This recall has been terminated (originally issued February 22, 2011).

Company
Haemonetics Corporation
Recall Initiated
January 25, 2011
Posted
February 22, 2011
Terminated
January 14, 2013
Recall Number
Z-1391-2011
Quantity
309 units
Firm Location
Braintree, MA
Official Source
View on FDA website ↗

Reason for Recall

Update Instructions for Use for the CardioPAT® Manual for Leak Detection. User may mis-interpret the air-leak rate displayed by the cardioPAT device when making a determination about chest-drain removal, which could result in the patient experiencing a pneumothorax.

Distribution

Worldwide distribution, including USA, Australia, Belgium, Canada, Switzerland, China, Germany, Finland, France, United Kingdom, Hong Kong, Israel, Italy, Japan, Netherlands, Russian Federation, and Sweden.

Lot / Code Info

All serial numbers

Root Cause

Labeling design

Action Taken

Haemonetics issued a Field Correction Notice on 1/25/11 . with the Air Leak Detection addendum and index be inserted into the appropriate document, applied to all cardioPAT@ devices on site. The serial number for each cardioPAT device on site should be recorded on the attached acknowledgement and the completed acknowledgement should be returned to your Haemonetics Sales Consultant. A Haemonetics Sales Consultant will train on the content of these items and insert them into the Operator Manuals