RecallDepth

ARTERIAL LINE KIT REF/REORDER NO. 498410B DISPOSABLE, SINGLE USE, Rx. Argon Medical Devices, Inc. 1445 Flat Creek Road Athens, TX 75751.

This recall has been terminated (originally issued June 2, 2010).

Company
Argon Medical Devices, Inc
Recall Initiated
March 29, 2010
Posted
June 2, 2010
Terminated
July 29, 2010
Recall Number
Z-1741-2010
Quantity
10 kits
Firm Location
Athens, TX
Official Source
View on FDA website ↗

Reason for Recall

There are 1-2 places on some of the multi-purpose kit blisters/trays in which there are holes and/or the material is so thin that holes are easily created. This would potentially cause the sterility of the products to be compromised.

Distribution

Distributed to United States (IL, AL, NY, TX, OK, MI, NC, CA, OH, VA, KY, CO, LA, WA, MO, AND IA) and Canada.

Lot / Code Info

Catalog Number: 498410B, Lot Number: 99737665, Expiration Date: 2012-10.

Root Cause

Package design/selection

Action Taken

Argon Medical Devices issued an "Urgent Medical Device Recall" notification via certified mail dated March 29, 2010 to distributors and end users. A Response Form was attached. Consignees were informed of the affected product and asked to quarantine and return all unused product to the firm. For further information, contact Argon Medical Devices at 1-903-677-9375.

More recalls by Argon Medical Devices, Inc

View all recalls by this company →