Cook PTFE Wire Guide with 3 cm flexible tip, to establish a tract for ureteral/renal access, .038", 145 cm; sterile; Order/Ref number 638413.
This recall has been terminated (originally issued August 9, 2007).
- Company
- Cook Urological, Inc.
- Recall Initiated
- June 12, 2007
- Posted
- August 9, 2007
- Terminated
- November 25, 2009
- Recall Number
- Z-1162-2007
- Quantity
- 100
- Firm Location
- Spencer, IN
- Official Source
- View on FDA website ↗
Reason for Recall
Lack of assurance of sterility, in that the product is labeled as sterile but the package may not have been sealed.
Distribution
Nationwide and Canada.
Lot / Code Info
Lot U1621121
Root Cause
Process control
Action Taken
Consignees were notified via letter dated 6/12/07 and instructed to cease using the product and to inform the firm if they have product to return.