CONTAK RENEWAL 3 CRT-D (model no. H170, H175, H177, H179). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. CONTAK RENEWAL 3 devices provide ventricular tachyarrhythmia and cardiac resynchronization therapies.

This recall has been terminated (originally issued May 23, 2006).

Company
Guidant Corporation
Recall Initiated
September 30, 2005
Posted
May 23, 2006
Terminated
January 16, 2007
Recall Number
Z-0890-06
Quantity
180
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Devices manufactured with batteries built on April 26 or 27, 2005, by a single operator, may have battery welds that may have been performed at settings outside of typical manufacturing specifications at the supplier.

Distribution

Nationwide (AL, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI). OUS to include: Australia, France, Germany, India, Ireland, Israel, Italy, Netherlands, Norway, South Africa, Spain, Switzerland, Turkey, United Kingdom.

Lot / Code Info

Model H170, serial numbers: 360449, 360632, 360655, 360657, 360661, 360666, 360714, 360719, 360721, 360738,  360741, 360742, 360743, 360745, 360746, 360749, 360756,  360758, 360763, 360766, 360790, 360791, 360792, 360796,  360800, 360802, 360844, 360847, 360848, 360861, 360863,  360904, 360907, 360908, 360910, 360986, 360987, 360988,  360989, 360991.  Model H175 serial numbers: 310338, 310434, 310523, 310532, 310533, 310537, 310544, 310545, 310706, 310707,  310709, 310713, 310719, 310721, 310734, 310735, 310737,  310746, 310753, 310796, 310798, 310813, 310907, 310908,  310909, 310913, 310916, 310976, 310990, 310996, 311007,  311010, 311011, 311013, 311015, 311016, 311092, 311096,  311152, 311154, 311181, 311185, 311386, 311389.   Model H177, serial numbers: 280992, 281011, 281012, 281017, 281020, 281021, 281022, 281024, 281045, 281056,  281066, 28107, 281077, 281119, 281121, 281123, 281124,  281126, 281127, 281143, 281175, 281213, 281233, 281247,  281320, 281356, 281364, 281368, 281371, 281376, 281385,  281401, 281403, 281413, 281476, 281480, 281484, 281485,  281491, 281533, 281539, 281542, 281544, 281559, 281562,  281580.   Model H179, serial numbers: 110905, 110907, 110908, 110950, 110952, 110954, 110955, 110958, 110980, 110983,  110987, 110988, 111000, 111001, 111003, 111004, 111007,  111012, 111024, 111041, 111042, 111128, 111173, 111180,  111198, 111201, 111248, 111250, 111254, 111263, 111294,  111299, 111308, 111313, 111334, 111335, 111336, 111338,  111340, 111341, 111342, 111346, 111350, 111352, 111390,  111391, 111392, 111393, 111397, 111422.

Root Cause

Other

Action Taken

Guidant issued a Physicain letter, dated 10/25/05 to implanting/following physicians. The letter states the issues and reommends patient follow-up care every three months, as stated in published labeling. An Advisory Update, dated 12/20/05, was sent to physicians that describes the issue, states that here have been zero reports of battery weld failures on implanted devices and the results of laboratory testing (accelerated testing) in which they tried to induce the failure. The results of the testing induced 2 battery weld post failures. A projected rate of occurence was also given. Patient management recommendations have not changed.