Lp(a) Diluent packaged into 60 mL bottles, Catalog #23817, Lot #s 238017-01 & 238017-03; 480 mL bulk catalog # 95046B, Lot # 238017-02. 60 mL bottles can be part of DiaSorin Lp(a) Antibody Reagent Kits Catalog # 86084, Lot #s 1144332, 114432A,13656, 113656A, and 113656B for use in DiaSorin Antibody Reagent Set for Lp(a) SPQ Test System .

Company
Diasorin Inc.
Recall Initiated
January 10, 2005
Posted
March 18, 2005
Terminated
July 22, 2006
Recall Number
Z-0636-05
Quantity
631 as individual vials, 1 bulk unit and 401 sold in kits
Firm Location
Stillwater, MN

Reason for Recall

QC testing of the Antibody Reagent Set for Lp(a) SPQ Test System it was discovered that the background was failing high of the specification. Through substitution testing it was determined that the Lp(a) diluent, part #238017 was the cause of the failing background specification.

Distribution

KC, MO, CA, IN, HI, GA, TX, MN, ME, AZ, FL, VA, MI, NJ, NC, NY, CO, WI, MS, BRUSSELS, GREECE, SWEDEN, ENGLAND, SPAIN, CANADA & SWITZERLAND

Lot / Code Info

Lp(a) Diluent - 60 mL bottles Catalog #23817, Lot #s 238017-01 & 238017-03; catalog # 95046B, Lot # 238017-02. DiaSorin Lp(a) Antibody Reagent Kits Catalog # 86084, Lot #s 1144332, 114432A,113656, 113656A, and 113656B

Root Cause

Other

Action Taken

Customers were sent notification letters and instructed to destroy any remaining inventory. Customers were also requested to return a form which indicates number of kits destroyed and a replacement was offered.