Lp(a) Diluent packaged into 60 mL bottles, Catalog #23817, Lot #s 238017-01 & 238017-03; 480 mL bulk catalog # 95046B, Lot # 238017-02. 60 mL bottles can be part of DiaSorin Lp(a) Antibody Reagent Kits Catalog # 86084, Lot #s 1144332, 114432A,13656, 113656A, and 113656B for use in DiaSorin Antibody Reagent Set for Lp(a) SPQ Test System .
- Company
- Diasorin Inc.
- Recall Initiated
- January 10, 2005
- Posted
- March 18, 2005
- Terminated
- July 22, 2006
- Recall Number
- Z-0636-05
- Quantity
- 631 as individual vials, 1 bulk unit and 401 sold in kits
- Firm Location
- Stillwater, MN
Reason for Recall
QC testing of the Antibody Reagent Set for Lp(a) SPQ Test System it was discovered that the background was failing high of the specification. Through substitution testing it was determined that the Lp(a) diluent, part #238017 was the cause of the failing background specification.
Distribution
KC, MO, CA, IN, HI, GA, TX, MN, ME, AZ, FL, VA, MI, NJ, NC, NY, CO, WI, MS, BRUSSELS, GREECE, SWEDEN, ENGLAND, SPAIN, CANADA & SWITZERLAND
Lot / Code Info
Lp(a) Diluent - 60 mL bottles Catalog #23817, Lot #s 238017-01 & 238017-03; catalog # 95046B, Lot # 238017-02. DiaSorin Lp(a) Antibody Reagent Kits Catalog # 86084, Lot #s 1144332, 114432A,113656, 113656A, and 113656B
Root Cause
Other
Action Taken
Customers were sent notification letters and instructed to destroy any remaining inventory. Customers were also requested to return a form which indicates number of kits destroyed and a replacement was offered.