Bucky Diagnost FS Part number 9890-010-83651
This recall has been terminated (originally issued December 24, 2003).
- Recall Initiated
- November 18, 2003
- Posted
- December 24, 2003
- Terminated
- April 11, 2005
- Recall Number
- Z-0268-04
- Quantity
- 3 units
- Firm Location
- Bothell, WA
- Official Source
- View on FDA website ↗
Reason for Recall
Support chain attachment on the Bucky carriage may fail and cause the counter weights or X-ray tube to fall.
Distribution
The firm distributed to three medical facilities in Florida, New York and California.
Lot / Code Info
Site numbers: 504495, 105355, 104447. The firm uses site numbers instead of serial numbers.
Root Cause
Other
Action Taken
The firm mailed a letter to their customers on November 18, 2003 stating upgrades will begin November 14, 2003.