One Cat. Life Pulse High Frequency Ventilator, Catalog Number: 203A, Manufactured by: Bunnell Incorporated Salt Lake City, Utah, Rx Only.

This recall has been terminated (originally issued January 14, 2011).

Company
Bunnell, Inc.
Recall Initiated
October 25, 2010
Posted
January 14, 2011
Terminated
April 17, 2012
Recall Number
Z-0901-2011
Quantity
660 total
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Product software coding error may compromise patient safety.

Distribution

Worldwide Distribution -- USA, Australia, Canada, Czech Republic, Korea, Malaysia, Trinidad & Tobago, and Uruguay.

Lot / Code Info

Model 203A

Root Cause

Software design

Action Taken

Customers were notified by an "Urgent: Medical Device Recall or Correction" letter starting 10/22/10. The letter identified the product and described the issues. Customers were given the option of either returning their ventilator(s) for rework or to receive installation/re-test instructions with an upgrade kit to rework themselves. The letter states that all Life Pulse ventilators in clinical use must be upgraded and there is no charge for the software upgrade.