RecallDepth

ABL90 FLEX Blood Analyzer Part Number: 393-090. Intended for use by trained technologists, nurses, physicians and therapists and for use in a laboratory environment, near patient or point-of-care setting. Also, Intended for in vitro testing of samples of heparinized whole blood.

This recall has been terminated (originally issued February 2, 2011).

Company
Radiometer America Inc
Recall Initiated
November 1, 2010
Posted
February 2, 2011
Terminated
May 9, 2012
Recall Number
Z-1011-2011
Quantity
432 units
Firm Location
Westlake, OH
Official Source
View on FDA website ↗

Reason for Recall

Due to a software error, a patient mix up can occur on the ABL90. If a sample is pre-registered and the sample "fails" the next result will inherit the patient data from the previous result.

Distribution

Worldwide Distribution -- USA, including states of MI, NJ, & PA and countries Australia, Austria, Belgium, Denmark, Finland, France, Germany, Indonesia, Italy, Japan, Kuwait, Malaysia, Netherlands, Norway, Poland, Russia, Spain, Sweden, Switzerland, United Arab Emirates & United Kingdom.

Lot / Code Info

S/N 393-090R0059N007, 393-090R0061N008 & 393-090R0061N009

Root Cause

Software design

Action Taken

On 11/16/10 the firm sent Medical Device Field Correction letters to their customers. The letter states that all systems will be upgraded and until each system is upgraded, they instructed their customer to "always query for patient ID prior to introducing the sample." They also instruct each customer to complete the "Fax Return Form" acknowledging receipt of the letter. If customers have questions regarding the letter or the upgrade, they should contact Technical Support at 1-800-736-0600 Opt 4.

More recalls by Radiometer America Inc

View all recalls by this company →